Mammomat 1000/3000
Register 3
SP B7-230.033.07
Page 2 of 12
Siemens-Elema AB
Installation and Start-Up
Rev. 03
11.99
SPS-UD
Solna, Sweden
11 - 2
Start-up and functional test of the IONTOMAT
Sensitivity correction
11
Indicates the sensitivity correction for different types of object tables. The position "without
grid" applies to Stereotactic Biopsy Attachment as well.
Factory set to "0", "0", "0".
Different film-density corrections can be programmed for the different tables with the
"Sens-cor" module.
Correction curve
11
Program "Correction curve" according to table 1 above (see also M3000 Service Instruc-
tions).
If the customer uses another combination or a mixture of those listed in table 1,
please refer to page 11 - 3.
Values shown as delivered.
Fig. 2
Fig. 3
MAM00187
IONTOMAT PM - Sens. correction
Without grid
1 Help
2 Save
3
4
5
6
7
8
9
10 Quit
With grid
Valid entries from -8 to 8 in 1/2 E.P.
For pos entries 1 st digit is < space >, for neg entries type < - >.
Input of decimal point not required!