11 - 1
Siemens-Elema AB
Register 3
SP B7-230.033.07
Page 1 of 12
Mammomat 1000/3000
Solna, Sweden
Rev. 03
11.99
SPS-UD
Installation and Start-Up
Start-up and functional test of the IONTOMAT
11
Checking and programming with the service PC
11
Use the service PC to check and program the IONTOMAT.
Operation of the service PC is described in Chapter 20 "Appendix".
The tables in this kit are optimized for the following four film/screen combinations.
Table 1
Program the IONTOMAT values as follows:
Sensitivity
11
Start with the sensitivity listed in the table for the film/screen combination used by the cus-
tomer. Program this values for "wing 1" and "wing 2" under H. If the customer uses one
more film/screen combination, program this value for "wing 1" and "wing 2" under D, oth-
erwise program the same values as under H.
If the film/screen combination used by the customer is not listed in table 1, start
with sensitivity 11.
Values from 0 to 31 can be entered.
If it is a M 3000 stand equipped with one "wing" only and counter-
weight, the programming for "wing 2" shall be omitted. If it is a
M 1000 stand, the programming is done for "base plate".
Be sure to use the same type of cassette, screen and film in each
object table as the customer uses.
Fig. 1
Dupont Microvision C
Agfa Mamoray MR6
Fuji UM-MA HC
Kodak Min-RE
Correction
curve
Sensi-
tivity
Film
Screen
Cassette
Kodak Min-R
Kodak Min-R2
11
A
Fuji HR Mammo Fine
Fuji EC-MA
9.5
B
Agfa MR Detail S
Agfa MR Detail S
7
C
Dupont Quantavision ME Dupont Quantavision 12
D
MAM00185
NOTICE
NOTICE
MAM00186
IONTOMAT PM - Sensitivity
Wing 1
11.0
1 Help
2 Save
3
4
5
6
7
8
9
10 Quit
11.0
H
D
Wing 2
11.0
11.0
Valid entries from 0 to 31 in 1/2 E.P.
Input of decimal point not required!