Mammomat 1000/3000
Register 3
SP B7-230.033.07
Page 6 of 14
Siemens-Elema AB
Installation and Start-Up
Rev. 03
11.99
SPS-UD
Solna, Sweden
5 - 6
Cable connections
Connecting the cable harness to the generator
5
Unit control cable X1
5
1.
Connect the unit control cable X1 to D700 X1 (1).
Filament cable X8
5
1.
Connect the filament cable X8 to D700 X8 (2).
Power supply cable X14
5
1.
Connect the power supply cable X14 to D711 X14 (3).
2.
Strain-relieve it on the side with a cable tie (4).
IONTOMAT signal cable X10 and X11
5
1.
Connect the IONTOMAT signal cable connectors, X10 and X11, to D700 X10 (5)
and X11 (6).
Required EMC measures are described on page 5 - 1 in this chap-
ter.
Fig. 6
Cable harness at the generator
The Iontomat signal cable consists of two cables laid together as
one cable with two connectors labeled X10 and X11 at the end of
the cable.