![Siemens MAMMOMAT 3000 Скачать руководство пользователя страница 70](http://html.mh-extra.com/html/siemens/mammomat-3000/mammomat-3000_installation-and-start-up-instructions-manual_380661070.webp)
Mammomat 1000/3000
Register 3
SP B7-230.033.07
Page 12 of 12
Siemens-Elema AB
Installation and Start-Up
Rev. 03
11.99
SPS-UD
Solna, Sweden
11 - 12
Start-up and functional test of the IONTOMAT
Output values of constancy testing (§16 Germany)
11
Determine the output values of the constancy testing at approx. 30 kV with the customer´s
own constancy test body (NORMI 7 test body etc.) and enter them in the form stapled to
the acceptance test certificate. Both tables must be checked
Procedure:
1.
On the control panel, select a value of approx. 30 kV and AEC.
2.
Place the test body on the object table flush with the chest wall side.
3.
Slide film cassette into the object table.
4.
Trigger the exposure and develop the film.
5.
Measure the resolution (A-D), density (1-4) and the over-radiation on the chest
wall side and enter in the form (steel balls).
6.
Read off the pen dosimeter and record the dose value.
Fig. 8
Plexiglass steps:
Woven steel wire:
1 = 40 mm
A = 100
µ
m
2 = 43 mm
B = 80
µ
m
3 = 49 mm
C = 63
µ
m
4 = 46 mm
D = 40
µ
m
NORMI 7
test body
Double detector
Radiographic system
Film cassette
Pen dosimeter
(Front view)
Pen dosimeter
A
B
C
D
1 2 3 4
(Top view)
Radiographic
system
NORMI 7
test body
Test arrangement
Steel balls