Mammomat 1000/3000
Register 3
SP B7-230.033.07
Page 2 of 4
Siemens-Elema AB
Installation and Start-Up
Rev. 03
11.99
SPS-UD
Solna, Sweden
13 - 2
Checking the swivel-arm system
4.
Set the projection angle to 60 degrees with presetting knob (2/Fig. 1).
The preset angle is displayed for approx 3 s.
5.
Press one of the buttons (1/Fig. 1).
The rotary motion of the swivel-arm system shall stop automatically when the
preset projection angle (both + and -) is reached and at 0 degrees.
Vertical adjustment
13
1.
Press the buttons for upward and downward movement (3/Fig. 1) of the swivel-
arm system.
The movement shall stop automatically when either end position is reached or
when the button is released.
Fig. 2
Display at the lower part of the stand
All eight rotary motion buttons (four on either side) must be
checked. After the button has been released, the swivel-arm sys-
tem shall remain in the position set.
Before vertical adjustment of the swivel-arm system is performed,
be sure to remove the protective strips from the metal curtain, see
page 3 - 5.
All eight vertical adjustment buttons (four on either side) must be
checked. The switching off in upper and lower end position is
effected by switches S881 and S884 respectively.
F
mm
o
OC
MA
M0
0216
NOTICE
CAUTION
NOTICE