12 - 1
Siemens-Elema AB
Register 3
SP B7-230.033.07
Page 1 of 2
Mammomat 1000/3000
Solna, Sweden
Rev. 03
11.99
SPS-UD
Installation and Start-Up
Further programming
12
Setting the real time clock
12
Prerequisite
12
The correct date and time must be set in the service PC. If not set date at C:\> by typing date
<ENTER> and time at C:\> by typing time <ENTER>.
Setting the time in the Mammomat
12
Main menu:
1.
Select Configuration <
↑
> <
↓
> < ENTER>.
2.
Select Clock <
↑
> <
↓
> < ENTER>.
3.
Press <F2>.
With <F2> the date and time set in the service PC are taken over by the
Mammomat.
Setting the tube-current reduction
12
In Mammomat, the tube-current reduction is factory-set to "ON", which means that the
tube current is reduced to approx. 50 mA during the first 100 ms of the exposure. In this
way, possible grid lines are avoided.
If the tube-current reduction is set to "OFF", grid lines may occur at short exposure times
(< 100 ms).
Main menu:
1.
Select Configuration <
↑
> <
↓
> < ENTER>.
2.
Select Miscellaneous <
↑
> <
↓
> < ENTER>.
3.
Select Tube mA red. <
↑
> <
↓
> < ENTER>.
Tube-current reduction (TCR) at short exposure times.
TCR switch is ON/OFF.
Exposure time with 50 mA tube current is 100 ms.
4.
Switch ON or OFF with <
↓
>.
5.
Save the programmed values with <F2>.
The tube-current reduction is effective only in Iontomat mode.
NOTE