Mammomat 1000/3000
Register 3
SP B7-230.033.07
Page 10 of 12
Siemens-Elema AB
Installation and Start-Up
Rev. 03
11.99
SPS-UD
Solna, Sweden
11 - 10
Start-up and functional test of the IONTOMAT
Checking the automatic transparency adaption and OPDOSE
(MAMMOMAT 3000 only)
11
The unit factory-set with the following values, which are to be regarded as recommenda-
tion. Changes can be done with the service PC program (Configuration, Miscellaneous,
Auto limits)
.
When making test exposures, follow the instructions on page 11 - 8 for positioning the
plexiglass plates and measuring the density.
Start with 60 mm plexi and compress. Make sure that the display indicates at least
60 mm. Make an exposure and measure the optical density. Repeat this procedure with
50 mm, 40 mm and 20 mm plexi.
Change density with the density setting buttons on the control panel as needed (the panel
programming switch must be ON, see Service PC program Configuration, Miscella-
neous, Panel programming). The optical density for the four different programs shall be
as near as possible to 1.5 O.D. (or the density preferred by the customer). Store with the
memory button.
Thickness
kV
Anode/filter
combination
Th
≥
60 mm
26
W / Rh
60 > Th
≥
45 mm
27
Mo / Rh
45 > Th
≥
30 mm
27
Mo / Mo
30 mm > Th
26
Mo / Mo
Fig. 5
Installations with filter disc Mo 0.030 / Rh 0.025 /Rh 0.050
Thickness
kV
Anode/filter
combination
Th
≥
60 mm
28
Mo / Rh
60 > Th
≥
45 mm
27
Mo / Rh
45 > Th
≥
30 mm
27
Mo / Mo
30 mm > Th
26
Mo / Mo
Fig. 6
Installations with filter disc Mo 0.030 / Rh 0.025
Both "wings" must be checked, if the stand is equipped with
pivoting object-table arm (Bucky selector).
NOTICE