Mammomat 1000/3000
Register 3
SP B7-230.033.07
Page 6 of 6
Siemens-Elema AB
Installation and Start-Up
Rev. 03
11.99
SPS-UD
Solna, Sweden
10 - 6
Checks with high voltage
*
) total exposure time dependent on programmed sensitivity
Fig. 10
Fig. 11
30 kV
1 V/T (1 T <=> 5 kV)
30 kV
1 V/T (1 T <=> 5 kV)
158 mA
1 V/T (1 T = 40 mA)
125 mA
1 V/T (1 T <=> 40 mA)
*
)
100 ms/t
*
)
100 ms/T
Fig. 12
Fig. 13
30 kV
1 V/T (1 T <=> 5 kV)
30 kV
1 V/T (1 T <=> 5 kV)
158 mA
1 V/T (1 T <=> 40 mA)
125 mA
1 V/T (1 T <=> 40 A)
*
)
100 ms/T)
*
)
100 ms/T
Tube P40 MoW, MAMMOMAT 3000
* tungsten anode 0.3
Tube P40 MoW, MAMMOMAT 1000/3000
molybdenum anode 0.3
MAM00316
30 kV Iontomat w. grid (P = 4.7 kW)
40 mm plexiglass
MAM00315
30 kV Iontomat w. grid (P = 3.75 kW)
40 mm plexiglass
MAM00317
30 kV Iontomat (P = 4.7 kW)
40 mm plexiglass
MAM00318
30 kV Iontomat (P = 3.75 kW)
40 mm plexiglass