Installation and Start-Up
Rev. 03
11.99
SPS-UD
Solna, Sweden
5 - 10
Cable connections
6.
Shield the decompression button cable to the stand chassis as shown in Fig. 12.
Connect to terminals 2 and 3 in terminal block X803 on the inside of the stand
chassis, see Fig. 12.
Main voltage connection
5
The equipment has been set to 400 V, 2-phase, in the factory (see test certificate).
Adjustment to the line frequency is not needed. If the mains voltage is not 400 V, 2-phase
at the installation site, a reconnection of the mains is necessary.
Should a reconnection for the mains voltage be necessary, this must be done at trans-
former T10 as well as p.c. board D711 (insertion or removal of fuses and labeling is spec-
ified in the wiring diagram, see Line Input page 5-3).
Change the rated voltage labels to the correct voltage (there are two labels on the gener-
ator and one on the stand). Labels are included in the installation material.
Fig. 12 Connecting the decompression button cable at the stand
For more details about changing mains, insertion or removal of
jumpers, fuses, labeling etc, see Wiring Diagram.