Checking the swivel-arm system
13 - 3
Mammomat 1000/3000
Register 3
SP B7-230.033.07
Page 3 of 4
Siemens-Elema AB
Installation and Start-Up
Rev. 03
11.99
SPS-UD
Solna, Sweden
Emergency stop
13
1.
Press the emergency stop button (4/Fig. 1).
The button shall latch and all motorized movements of the equipment shall be
blocked ("Er 824" shall appear). To reset the emergency-stop button, turn it
clockwise.