13 - 1
Mammomat 1000/3000
Register 3
SP B7-230.033.07
Page 1 of 4
Siemens-Elema AB
Installation and Start-Up
Rev. 03
11.99
SPS-UD
Solna, Sweden
Checking the swivel-arm system
13
Rotary motion
13
1.
System "OFF".
The rotary motion and the vertical adjustment of the swivel-arm system shall be
blocked.
2.
System "ON", object table installed and projection angle set to minimum with
presetting knob (2/Fig. 1).
3.
Check the rotary motion of the swivel-arm system by pressing buttons (1/Fig. 1)
one by one.
The rotation of the swivel-arm system shall stop when the button is released or
when end position is reached (+135 degrees clockwise and -180 degrees
counterclockwise). The projection angle is shown on the display at the lower part
of the stand.