InterRanger DPS300
94
7ML19981BC01
•
after replacing the DPS300 EPROM with a different software revision number.
Note:
Following a Global Master Reset, complete reprogramming is required.
A
Point Number Master Reset
(reset the non-global parameters of the displayed
Point Number to preset values) should be performed if a Point Number is being
reassigned to a different clarifier.
To perform a Global Master Reset:
1. With the Parameter Number field accessed, key in 999,
2. With the Point Number field accessed, key in 00,
3. Press “C.ALL” is displayed until the reset is complete.
To perform a Point Number Master Reset:
1. With the Parameter Number field accessed, key in 999,
2. With the Point Number field accessed, key in the Point Number to be reset (1, or 2)
3. Press , “C.ALL” is displayed until the reset is complete.