InterRanger DPS300
84
7ML19981BC01
To select the desired Scope Display...
1. Press to display the Auxiliary Function symbol,
2. Press or to scroll access the desired Reading display symbols.
3. Press to select the displayed symbols.
Alternatively, a 4 digit binary value may be entered, where a “0” turns the
associated signal display “off”, and a “1” turns the associated display “on”.
e.g.
1110 = PCn_ = Echo Profile, TVT Curve, and Echo Marker displays on.
See TECHNICAL REFERENCE / Echo Processing for more Scope Display
illustrations.
Use the Scope Displays after pressing
(in the program mode) to observe the
result of Echo Processing parameter alterations. (Take several measurements to
verify measurement repeatability and overcome Echo Lock (P711) restrictions).
As preset, all displays are off.
P815 (V) Echo Time
Observe the time (in ms) between the transmit pulse and the processed echo,
including delays such as filtering and rate limit.
P816 (V) Echo Time Raw
Observe the time of flight (in ms) between the transmit pulse and the raw echo,
i.e. before processing.
Values:
0.000 to 9999
When a Profile Pointer Parameter is accessed, the Echo Lock Window scope
display changes to a Profile Pointer display. The Profile Pointer may be moved to
a number of points on the Echo Profile, to gain specific information dependent
upon the Profile Pointer Parameter used.
To move the Profile Pointer to a specific point, enter the desired value. The Profile
Pointer will move to the nearest acceptable Echo Profile point. The Profile Pointer
is preset to “0”.
Alternatively, to scroll the Profile Pointer along the Echo Profile...
1. Press to display the Auxiliary Function symbol,
2. Press or to move the Profile Pointer to the left or right respectively.
When the Profile Pointer Parameters are exited and
is pressed or the RUN
mode is entered, the Profile Pointer display automatically changes back to the
Echo Lock Window display.
P817 (V) Profile Pointer Time
Observe the time (in ms) between the transmit pulse and the Profile Pointer.
Values:
0.000 to 9999