Transducer Mounting
Procure a length of corrosion resistant conduit for suspending the transducer*
such that the transducer face will be immersed 3 to 6 cm into the clarifier.
Siemens suggests ¾” standard stainless steel pipe (schedule 10) cut with ¾” NTP
threads.
* transducer or transducer and skimmer guard assembly. Refer to the Skimmer
Guard instructions (above).
Note:
If skimmer is in operation, shut off until mounting is complete
With the extension arms swung in towards the rail, clamp the conduit extension of
the transducer assembly to the extension arm ends and snug tight using the two
smaller u-bolts. Slide the conduit down so that the transducer face is immersed 3 to
6 cm into the clarifier water.
Note:
Run the transducer cable from the rigid conduit end, through a flexible conduit to
the junction box (if required) in a manner that allows the extension arm to swing
out fully.
3 to 6 cm