InterRanger DPS300
68
7ML19981BC01
ENHANCEMENT PARAMETERS
This section identifies all DPS300 operator programmable features designed for
altering operation to suit individual operator preferences or overcome
measurement difficulties.
Typically, these parameters are only altered as directed by the
TROUBLESHOOTING GUIDE. If RUN mode performance requires improvement,
access and alter the following parameters as required.
Note:
Otherwise, proceed to OPERATION.
PARAMETER CATEGORY
PURPOSE
PAGE
Data Logging
to view previous maximum Temperature records
68
Profile Records
intended for use by Siemens Service Personnel
69
Installation Records
to identify length of service and power failure occurrence
72
Range Calibration
to compensate for measurement offset and/or sound velocity
72
Temperature Compensation
to override automatic temperature compensation features
74
Rate
to adjust the rate of response to change in the sludge level
75
Measurement Verification
to adjust the rate of response to change in the sludge level
77
Display Parameters
to override the preset display operating characteristics
78
Echo Processing
to troubleshoot false echo detection
81
Advanced Echo Processing
intended for use by Siemens Service personnel
83
Test Parameters
intended for use by Siemens Service personnel
90
Measurement Parameters
to verify Application Parameter programming
92
Master Reset
to reset Parameter Values to factory settings
93
Note:
When an Enhancement Parameter is altered, return to the RUN mode to verify the
performance desired is achieved, before making any other changes.
Data Logging Parameters (P300 and P302)
View the maximum Temperature recorded in the RUN mode.
P300 (V) Temperature, Transducer Max
Use this feature to view the highest temperature recorded (in ° C) by the XCT-12
transducer.
Values:
-50 to 150
P302 (V) Temperature, Sensor Max
Use this feature to view the highest temperature recorded (in ° C) by the TS-3
temperature sensor, if used.
When accessed, the TS-3 symbol is displayed in the Point Type field and the TS-3
Number is displayed in the Point Number field.
Values:
-50 to 150