5-18 Image Optimization
4. Press <PW> again or <Update> to enter PW mode and perform the examination. You can
also adjust the SV size, angle and depth in real-time scanning.
5. Adjust the image parameters during PW mode scanning to obtain optimized image.
6. Perform other operations (e.g. measurement and calculation) if necessary.
5.8.2
PW Mode Image Parameters
In PW mode scanning, the image parameter area in the upper right corner of the screen
displays the real-time parameter values as follows:
Parameter F
G
PRF WF
SVD
SV
Angle
Meaning
Frequency
Gain
PRF
Wall Filter
SV Position
SV Size
Angle
When you adjust the depth of the B mode image, related changes will occur in PW mode
image as well.
During PW mode imaging, you can switch between B and PW menu from the menu title;
for details, please refer to “5.1 Switching Among Imaging Modes” chapter.
5.8.3
PW Mode Image Optimization
Gain
Description
This function is intended to adjust the gain of spectrum map. The real-time
gain value is displayed in the image parameter area in the upper right corner
of the screen.
Operations
Rotate the [Gain/iTouch] knob to adjust the gain.
The adjusting range is 0-100.
Effects
Increasing the gain will brighten the image and you can see more received
signals. However, noise may also be increased.
Frequency
Description
Refers to the operating frequency in PW mode of the probe, the real-time value
of which is displayed in the image parameter area in the upper left corner of
the screen.
Operation
Select the frequency value through the [Frequency] item in the image menu or
rotate the <Focus/Freq./THI> knob on the control panel.
Values of frequency vary depending upon the probe types.
Select the frequency according to the detection depth and current tissue
characteristics.
Effects
The higher the frequency, the better the resolution and sensitivity, and the
worse the force of penetration.
Baseline
Description
Refers to the area where the velocity is zero in the spectrum.
Operations
Press the <Baseline> key on the control panel (the <Baseline> key backlight is
on) and rotate the Multifunctional Knob on the right to adjust.
Effects
To change the range of flow velocity to optimize the image.
PW Steer
Description
This function provides adjustment on angles for sampling line.
Содержание DC-N2
Страница 1: ...DC N2 DC N2T DC N2S Diagnostic Ultrasound System Operator s Manual Basic Volume...
Страница 2: ......
Страница 14: ......
Страница 26: ......
Страница 56: ......
Страница 66: ......
Страница 106: ......
Страница 114: ...6 8 Display Cine Review 6 5 Preset Open Setup System Preset General to preset the cine storage length Clip length 1 60s...
Страница 142: ......
Страница 168: ......
Страница 206: ......
Страница 216: ......
Страница 221: ...P N 046 005132 00 V4 0...