Safety Precautions 1-3
1.4 Safety
Precautions
Please observe the following precautions to ensure patient and operator’s safety when using
this system.
DANGER:
DO NOT use flammable gasses, such as anesthetic gas or
hydrogen, or flammable liquids such as ethanol, near this
system, because there is danger of explosion.
WARNING:
1.
Do connect the power plug of this system and power plugs
of the peripherals to wall receptacles that meet the ratings
indicated on the rating nameplate. Using a multifunctional
receptacle may affect the system protective grounding
performance, and cause the leakage current to exceed
safety requirements.
Use the cable provided with this system to connect the
printer. Other cables may result in electric shock.
You must use the power adapter provided with the system;
otherwise electric shock may result.
You can only adopt the power supply method provided by
Mindray, other power supply modes (e.g. using a UPS) may
result in electric shock.
2.
Connect the protective grounding conductor before
turning ON the system. Disconnect the grounding cable
after turning OFF the system. Otherwise, electric shock
may result.
3.
For the connection of power and grounding, follow the
appropriate procedures described in this operator’s
manual. Otherwise, there is risk of electric shock. DO
NOT connect the grounding cable to a gas pipe or water
pipe; otherwise, improper grounding may result or a gas
explosion may occur.
4.
Before cleaning the system, disconnect the power cord
from the outlet. System failure and electric shock may
result.
5.
This system is not water-proof designed. Do Not use this
system in any place where water or any liquid leakage
may occur. If any water is sprayed on or into the system,
electric shock
or device malfunction
may result. If water is
accidentally sprayed on or into the system, contact
Mindray Customer Service Department or sales
representative.
6.
Do not use a transducer that has a damaged, scratched
surface, or exposed wiring of any kind. Immediately stop
using the transducer and contact Mindray Customer
Service Department or sales representative. There is risk
of electric shock if using a damaged or scratched
transducer.
Содержание DC-N2
Страница 1: ...DC N2 DC N2T DC N2S Diagnostic Ultrasound System Operator s Manual Basic Volume...
Страница 2: ......
Страница 14: ......
Страница 26: ......
Страница 56: ......
Страница 66: ......
Страница 106: ......
Страница 114: ...6 8 Display Cine Review 6 5 Preset Open Setup System Preset General to preset the cine storage length Clip length 1 60s...
Страница 142: ......
Страница 168: ......
Страница 206: ......
Страница 216: ......
Страница 221: ...P N 046 005132 00 V4 0...