Image Optimization 5-11
Effects
Adjust according to the actual situation and obtain a desired analysis through
comparison.
Speed
Description
This function is used to set the scanning speed of M mode imaging, and the
real-time speed value is displayed in the image parameter area in the upper
left corner of the screen.
Operation
Change the speed through the [Speed] item on the menu;
There are 6 levels of scan speed available, the smaller the value the faster the
speed.
Effects
Speed changing makes it easier to identify disorders in cardiac cycles.
Tint Map
Description
Colorize function provides an imaging process based on color difference
rather than gray distinction.
Operation
Select the colorize map through the [Colorize Map] item on the menu;
The system provides 16 different maps to be selected among.
Impacts
The function is available in real-time imaging, freeze or cine review status.
Gray Map
Description
Adjusting grayscale contras to optimize the image.
Operation
Select among the maps through the [Gray Map] item on the menu.
The system provides 8 gray maps to be selected among.
Impacts
The function is available in real-time imaging, freeze or cine review status.
Edge Enhance
Description
This function is used to increase image profile, so as to distinguish the image
boundary.
Operation
Adjust through the [Edge Enhance] item on the menu;
The system provides 14 levels of edge enhance effects, off represents no
edge enhance is turned on, and the bigger the value the stronger the effect.
Impacts
Larger edge enhance may lead to noise increasing.
Dynamic Range
Description
Adjusts contrast resolution of an image, compresses or expands gray display
range. The real-time dynamic range value is displayed on the image
parameter area in the upper left corner of the screen.
Operation
Adjust through the [Dynamic Range] item on the menu.
The adjusting range is 30-220 in increments of 5.
Effects
The more the dynamic range, the more specific the information, and the lower
the contrast with more noise.
Содержание DC-N2
Страница 1: ...DC N2 DC N2T DC N2S Diagnostic Ultrasound System Operator s Manual Basic Volume...
Страница 2: ......
Страница 14: ......
Страница 26: ......
Страница 56: ......
Страница 66: ......
Страница 106: ......
Страница 114: ...6 8 Display Cine Review 6 5 Preset Open Setup System Preset General to preset the cine storage length Clip length 1 60s...
Страница 142: ......
Страница 168: ......
Страница 206: ......
Страница 216: ......
Страница 221: ...P N 046 005132 00 V4 0...