DICOM 10-13
In the off-line status, you can:
Perform a second query; or,
Click [Show Detail] to view the patient information in details.
10.3.4 MPPS
MPPS is used to send exam state information to the configured server. This will facilitate the
other systems to obtain the exam progress in time.
The status information is described as below:
z
When you begin an exam or send image(s) during the exam, the system sends
status information “Active” to MPPS server.
z
When the exam is completed, the system sends status information “End” to MPPS
server.
z
When a paused exam is continued, the system sends status information “Active” to
MPPS server.
z
When an exam is cancelled, the system sends status information “Cancelled” to
MPPS server.
10.3.5 Storage
Commitment
Storage commitment is used to make sure the images are successfully stored in the DICOM
server.
Before storage commitment, you should set a default storage commitment server.
Storage commitment after sending images on iStation screen
(1) Open iStation screen: press <iStation> or press <Patient> and click [iStation].
(2) Select an exam record (image (s) is/are stored in the exam record), click
to
open the Send To dialogue box.
(3) Click to select “DICOM” in the “Target” box on the left side, then select the DICOM
storage server in the “Storage Server” box on the right side.
(4) Click [OK] to start sending. The system will send all the images stored in the exam
record to the storage server, meanwhile, it will send storage commitment to the
storage commitment server.
To send storage commitment after an exam ends
(1) Open [Setup] -> [System Preset]-> [General], and then select [Sending/ Printing after
End Exam] in the Patient Info area.
(2) Set the default storage server and storage commitment server.
a) Enter the DICOM Service Preset screen via “[Setup] -> [DICOM Preset] -> [Set
DICOM Service]”.
b) Select a server in the Service List of Storage page and Storage Commitment
page and click [Default]; you can see “Y” is marked in the Default column.
c) Click [Exit] to exit the page and return to Setup menu, then click [Save] on the
Setup menu to make the preset to take effect.
(3) After performing the above settings, each time you press <End Exam> on the control
panel, the system will send the image to the default DICOM storage server for
storage and send storage commitment to storage commitment server.
If images are successfully sent to the storage server, the storage commitment server will
return information about the successful image storage. In the iStation screen, you will see
there is a tick “
√
” marked in the list below
.
Tips:
Storage commitment is confined to the whole exam; not each image sending can be indicated.
Содержание DC-N2
Страница 1: ...DC N2 DC N2T DC N2S Diagnostic Ultrasound System Operator s Manual Basic Volume...
Страница 2: ......
Страница 14: ......
Страница 26: ......
Страница 56: ......
Страница 66: ......
Страница 106: ......
Страница 114: ...6 8 Display Cine Review 6 5 Preset Open Setup System Preset General to preset the cine storage length Clip length 1 60s...
Страница 142: ......
Страница 168: ......
Страница 206: ......
Страница 216: ......
Страница 221: ...P N 046 005132 00 V4 0...