![Shenzhen Mindray Bio-Medical Electronics Co., Ltd. DC-N2 Скачать руководство пользователя страница 218](http://html1.mh-extra.com/html/shenzhen-mindray-bio-medical-electronics-co-ltd/dc-n2/dc-n2_operators-manual_1256057218.webp)
B-2 Electrical Safety Inspection
B.2.1 Visual
Inspection
Test Item
Acceptance Criteria
The enclosure and
accessories
No physical damage to the enclosure and accessories.
No physical damage to meters, switches, connectors, etc.
No residue of fluid spillage (e.g., water, coffee, chemicals,
etc.).
No loose or missing parts (e.g., knobs, dials, terminals,
etc.).
B.2.2 Contextual
Inspection
Test Item
Acceptance Criteria
The enclosure and
accessories
No unusual noises (e.g., a rattle inside the case).
No unusual smells (e.g., burning or smoky smells, particularly
from ventilation holes).
No taped notes that may suggest device deficiencies or
operator concerns.
B.3 Device
Labeling
Check the labels provided by the manufacturer or the healthcare facility are present and
legible.
z
Main unit label
z
Integrated warning labels
B.4 Protective
Earth
Resistance
a. Plug the probes of the analyzer into the device’s protective earth terminal and
protective earth terminal of the AC power cord.
b. Test the earth resistance with a current of 25 A.
c. Verify the resistance is less than limits.
LIMITS
ALL COUNTRIES R = 0.2
Ω
Maximum
B.5 Earth
Leakage
Test
Run an Earth Leakage test on the device being tested before performing any other
leakage tests.
The following outlet conditions apply when performing the Earth Leakage test.
z
normal polarity( Normal Condition),
z
reverse polarity( Normal Condition),
z
normal polarity with open neutral(Single Fault Condition),
z
reverse polarity with open neutral(Single Fault Condition)
Содержание DC-N2
Страница 1: ...DC N2 DC N2T DC N2S Diagnostic Ultrasound System Operator s Manual Basic Volume...
Страница 2: ......
Страница 14: ......
Страница 26: ......
Страница 56: ......
Страница 66: ......
Страница 106: ......
Страница 114: ...6 8 Display Cine Review 6 5 Preset Open Setup System Preset General to preset the cine storage length Clip length 1 60s...
Страница 142: ......
Страница 168: ......
Страница 206: ......
Страница 216: ......
Страница 221: ...P N 046 005132 00 V4 0...