![Shenzhen Mindray Bio-Medical Electronics Co., Ltd. DC-N2 Скачать руководство пользователя страница 76](http://html1.mh-extra.com/html/shenzhen-mindray-bio-medical-electronics-co-ltd/dc-n2/dc-n2_operators-manual_1256057076.webp)
5-10 Image Optimization
5.5 M
Mode
5.5.1
M Mode Exam Protocol
1. Select a high-quality image during B mode scanning, and adjust to place the area of
interest in the center of the B mode image.
2. Press <M> on the control panel, and roll the trackball to adjust the sampling line.
3. Press <M> on the control panel again or <Update> to enter M mode, then you can
observe the tissue motion along with anatomical images of B mode.
4. During the scanning process, you can also adjust the sampling line accordingly when
necessary.
5. Adjust the image parameters to obtain optimized images.
6. Perform other operations (e.g. measurement and calculation) if necessary.
5.5.2
M Mode Parameters
In M mode scanning, the image parameter area in the upper left corner of the screen
displays the real-time parameter values as follows:
Display
F
D
DR
G
V
Parameter
Frequency Depth
M Dynamic Range
M Gain
M
speed
During M mode imaging, you can switch between B and M menu from the menu title; for
details, please refer to “5.1 Switching Among Imaging Modes” chapter.
During M Mode scanning, frequency and acoustic power of the transducer are
synchronous with that of B Mode.
Adjustment of the depth or TGC to the B Mode image will lead to corresponding changes
in M Mode image.
5.5.3
M Mode Image Optimization
Gain
Description
To adjust the gain of M mode image. The real-time gain value is displayed in
the image parameter area in the upper left corner of the screen.
Operation
Rotate the <Gain/ iTouch> knob clockwise to increase the gain, and anti-
clockwise to decrease.
The adjusting range is 0-100.
Effects
Increasing the gain will brighten the image and you can see more received
signals. However, noise may also be increased.
Focus Position
Description
To change the focus position in M mode, symbols as " " of which are
displayed on the right of the image.
Operation
Adjust the focus position through the [Focus Position] item on the menu.
Display Format
Description
To set the display format of M mode image with B mode image.
Operation
Adjust through the [Display Format] item on the menu;
There are 4 formats available for the images display: V1:1, V1:2, V2:1, Full.
Содержание DC-N2
Страница 1: ...DC N2 DC N2T DC N2S Diagnostic Ultrasound System Operator s Manual Basic Volume...
Страница 2: ......
Страница 14: ......
Страница 26: ......
Страница 56: ......
Страница 66: ......
Страница 106: ......
Страница 114: ...6 8 Display Cine Review 6 5 Preset Open Setup System Preset General to preset the cine storage length Clip length 1 60s...
Страница 142: ......
Страница 168: ......
Страница 206: ......
Страница 216: ......
Страница 221: ...P N 046 005132 00 V4 0...