15-2 Guidance and Manufacturer’s Declaration
TABLE 1
GUIDANCE AND MINDRAY DECLARATION—ELECTROMAGNETIC EMISSIONS
The system is intended for use in the electromagnetic environment specified below. The
customer or the user of system should assure that it is used in such an environment.
EMISSIONS TEST
COMPLIANCE
ELECTROMAGNETIC ENVIROMENT
-
GUIDANCE
RF emissions
CISPR 11
Group 1
The system
uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not likely
to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The system
is suitable for use in all
establishments including domestic
establishments and those directly
connected to the public low-voltage power
supply network that supplies buildings
used for domestic purposes
Harmonic Emissions
IEC 61000-3-2
Class A
Voltage Fluctuations/ Flicker
Emissions
IEC 61000-3-3
Compliance
Содержание DC-N2
Страница 1: ...DC N2 DC N2T DC N2S Diagnostic Ultrasound System Operator s Manual Basic Volume...
Страница 2: ......
Страница 14: ......
Страница 26: ......
Страница 56: ......
Страница 66: ......
Страница 106: ......
Страница 114: ...6 8 Display Cine Review 6 5 Preset Open Setup System Preset General to preset the cine storage length Clip length 1 60s...
Страница 142: ......
Страница 168: ......
Страница 206: ......
Страница 216: ......
Страница 221: ...P N 046 005132 00 V4 0...