01910001B_8
© Servomex Group Limited. 2020
83
3.
Fully �ghten the cover. If necessary, fit a suitable metal bar between the protruding fins
on the front of the cover to provide addi�onal
leverage.
4.
Use the 2 mm hexagonal key supplied to �ghten the locking screw which secures the
cover.
5.
Ensure
that the cer�fica�on label is in the correct orienta�on. If necessary, press and
turn the label un�l it is in the correct orienta�on.
Note:
If a new fuse fails immediately after you have fitted it, there may be an electrical fault
in the analyser: contact Servomex or your Servomex agent.
8.5
Use of the analyser
with toxic
gases
If you use the analyser to sample toxic gases, the concentra�ons of the gases
sampled or used for calibra�on of the analyser may be above their respec�ve
threshold limit values. You must therefore regularly leak
-
test the analyser and
associated equipment. If any leaks are found, do not con�nue to use the analyser or
associated equipment un�l the leaks have been sealed.
When you carry out
a leak test, do not exceed a maximum pressure of 34.5 kPa gauge
(0.35 bar gauge, 5 psig) and do not introduce a sudden change of pressure into the
analyser. If you do, you may damage it.
If you use a liquid to assist in leak tes�ng, do not spill liquid onto any of the electrical
components in the sample/control compartment.
If you use the analyser
for toxic gas sample measurements, you must regularly leak
- test the
analyser
and the associated sample inlet and outlet pipelines or hoses.
We recommend that you leak
-test the
analyser
at least once every 6 months:
•
If
there are leaks within the analyser, it must be returned to Servomex for repair. Do not
con�nue to use the analyser
.
•
You must seal any leaks in your sample pipelines or
system.
Note:
When you leak
-
test, ensure that you do not exceed the maximum pressure, and do not
increase the pressure in the analyser too quickly (see the caution above): we
recommend that you allow at least 30 seconds to fully pressurise the analyser to the
maximum press
ure.
8.6
Preventative
maintenance
To minimise unscheduled analyser
down�me, ensure the proper opera�on of the analyser
and to comply
with the guidelines of applicable regulatory bodies, we recommend that you u�lise the SERVOSURE
annual preventa�ve maintenance program for your analyser
.
Содержание SERVOTOUGH Oxy
Страница 2: ...This page inten onally blank...
Страница 4: ...This page inten onally blank...
Страница 20: ...8 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 28: ...16 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 30: ...18 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 41: ...01910001B_8 Servomex Group Limited 2020 29 Figure 12 Fixing dimensions mm...
Страница 82: ...70 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 98: ...86 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 110: ...98 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 112: ...100 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 126: ...114 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 130: ...118 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 132: ...120 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 140: ...128 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 142: ...130 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 144: ...132 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...