Ensure that the cables that you connect to the analyser
are routed so that they do not
present a trip hazard.
•
When you carry out insula�on tes�ng, disconnect all cables from the
analyser
.
5.3.2
Configure the cable entry
holes
If the analyser
is installed in a hazardous area, all cable
-
glands, adaptors and blanking
plugs must be suitably approved.
•
ATEX approved
analysers must be fited with IP66 ATEX Ex d IIC approved
components.
•
IECEx approved
analysers must be fited with IP66 IECEx Ex d IIC
approved
components.
•
CSA North American Approved analysers must be fited with CSA/UL, cULus
or cCSAus 4X, Class I, Groups A, B, C, D approved components.
The
analyser
has two 3/4 inch and two 1/2 inch NPT cable entry holes. As supplied, three of these holes
are fited with blanking plugs (rated to IP66 / NEMA 4X).
Configure
the
analyser
so that the entry holes that you use for the electrical supply and signal
connec�on cables have suitable cable
-
glands (and/or adaptors) fited, and so that any unused entry hole
has a blanking plug fited.
All cable
-
glands/adaptors and plugs used must provide ingress protec�on to IP66, NEMA 4X or beter.
Note:
It is recommended that the 1/2 inch NPT cable entry hole on the far right hand side of
the enclosure is used for the power supply cable.
5.3.3
Interface signal
connections
All of the analyser
interface signal outputs are considered to be incendive and
therefore must only be connected to safe area equipment.
To comply with EMC requirements, you must use a screened 4 to 20 mA cable to
connect to the mA output. It is recommended that screens
are terminated at the
cable gland.
Connect the wires in your cables to the appropriate terminals as described below and refer to Sec�on
for detailed ra�ngs and cable sizes.
1.
Pass your signal cable(s) through one or two
of the cable
-
glands fited to the base of the
power/interface compartment.
2.
Refer to Figure 4. Connect the mA output wires in your cables to the corresponding
terminal pins of the mA output, as follows:
Содержание SERVOTOUGH Oxy
Страница 2: ...This page inten onally blank...
Страница 4: ...This page inten onally blank...
Страница 20: ...8 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 28: ...16 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 30: ...18 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 41: ...01910001B_8 Servomex Group Limited 2020 29 Figure 12 Fixing dimensions mm...
Страница 82: ...70 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 98: ...86 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 110: ...98 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 112: ...100 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 126: ...114 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 130: ...118 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 132: ...120 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 140: ...128 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 142: ...130 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 144: ...132 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...