![Servomex SERVOTOUGH Oxy Скачать руководство пользователя страница 79](http://html1.mh-extra.com/html/servomex/servotough-oxy/servotough-oxy_operators-manual_1247065079.webp)
01910001B_8
© Servomex Group Limited. 2020
67
Figure
67 -
Filter factor screen
3.
Edit the filter factor to the desired level. Values must be between 001, no filtering, and
064.
Figure
68 -
Filter Factor Edit Screen
Note:
The value relates to the number of rea
dings that the filter averages over. i.e. 15 would
mean a filter length of 15 readings. The measurement reading updates five times per
second, therefore in this case the filter length is three seconds of measurement data.
Note:
It is only
recommended that the filter level is altered, default value of 8 readings, if
excessive measurement fluctuations are experienced. This could be caused by high
levels of vibration or sample system fluctuations, and we recommend that the root
causes of the fluctuations are fully investigated before applying higher filter levels, as
increasing filter length increases the response time of the
analyser
i.e. at a filter
length of 064 the response time (T90) of the
analyser
is approximately 19 seconds.
5.21
Setting up
the Flowcube (F
3
) Flow Sensor
(option)
The
flow sensor op�on measures the sample flow through a measurement transducer and allows the
user to configure two low flow alarms and one high flow alarm. The flow sensor allows preventa�ve
maintenance to be taken before the sample flow rate reduces to unacceptable levels Sec�on 2.4.
Configura�on op�ons and defaults for each flow sensor alarm level are given below:
Содержание SERVOTOUGH Oxy
Страница 2: ...This page inten onally blank...
Страница 4: ...This page inten onally blank...
Страница 20: ...8 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 28: ...16 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 30: ...18 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 41: ...01910001B_8 Servomex Group Limited 2020 29 Figure 12 Fixing dimensions mm...
Страница 82: ...70 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 98: ...86 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 110: ...98 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 112: ...100 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 126: ...114 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 130: ...118 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 132: ...120 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 140: ...128 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 142: ...130 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 144: ...132 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...