![Servomex SERVOTOUGH Oxy Скачать руководство пользователя страница 93](http://html1.mh-extra.com/html/servomex/servotough-oxy/servotough-oxy_operators-manual_1247065093.webp)
01910001B_8
© Servomex Group Limited. 2020
81
8.
ROUTINE
MAINTENANCE
The
analyser
must be maintained by a suitable skilled and competent person.
Sample and calibra�on gases may be toxic, asphyxiant or flammable.
Never inspect
the inlet filter, open the front cover of the sample/control compartment, or service or
repair the
analyser
while such gases are s�ll connected to it.
Sample and calibra�on gases may be toxic, asphyxiant or flammable and hazardous
concentra�ons may accumulate within the analyser
during use.
Always o
pen the
front cover of the sample/control compartment in a
force
-
ven�lated enclosure with a
minimum volume of 17 m3, o
r in another
appropriate environment, in which any
hazardous gases are directed away from you.
8.1
Cleaning the
analyser
When necessary, use a damp (but not wet) cloth to wipe clean the outer surfaces of the
analyser, to
prevent the entry of dust or other par�culates into the breather vent or the interior of the analyser
.
8.2
Inspection
It
is recommended that the analyser
is inspected for signs of corrosion. Par�cular care is should be taken
when inspec�ng the door seals, sample inlet and outlet entries, moun�ng plate and the gland entries.
If
a sample system or corrosive purge is being used, then the opera�on of these should be verified at
regular intervals, no greater than annually. Failure of either the sample system or the corrosive purge
can result in significant analyser
damage and research has shown that sample system failures account
for over 70% of analy�cal measurement field failures.
The pressure drop across the analyser
is low, less than 1.0kPa (0.15psi) in most cases, if excessive
pressure is required to generate a flow through the analyser
it is an indica�on that the inlet sintered
filter is blocked or contaminated. If this is the case,
then this fi�ng can be removed and inspected or
replaced.
8.3
Inspecting/replacing the fuse (when
necessary)
E
nsure that the electrical supply is isolated/locked out from the analyser. If you do
not, there will be a danger of injury or death from electric shock.
Do not remove the cover of the power/signal compartment if an explosive
atmosphere may be
present.
The power/interface compartment cover is heavy. Ensure that you do not drop the
cover once it is disengaged
from the body of the analyser. If you do, you may injure
Содержание SERVOTOUGH Oxy
Страница 2: ...This page inten onally blank...
Страница 4: ...This page inten onally blank...
Страница 20: ...8 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 28: ...16 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 30: ...18 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 41: ...01910001B_8 Servomex Group Limited 2020 29 Figure 12 Fixing dimensions mm...
Страница 82: ...70 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 98: ...86 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 110: ...98 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 112: ...100 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 126: ...114 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 130: ...118 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 132: ...120 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 140: ...128 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 142: ...130 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 144: ...132 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...