Relay
Relay coil state in
"active" status
Definition of "active" status
Fault
Deenergised
Measurement or
analyser
fault detected
Alarm 1
Energised
Alarm 1 ac�ve
Alarm 2
Energised
Alarm 2 ac�ve
mA Range Indica�on
Energised
mA
range has changed over from range 1
to range 2
Service in Progress (SIP)
Energised
Service in progress status enabled
Maintenance Required
Energised
Measurement or
analyser
maintenance
required detected
The relay coil status, energised or deenergised, can be set up to meet any par�cular applica�on
requirements.
Note:
Example:
As default, the fault relay coil is deenergised during an "active" status. As defined
above, this is when a measurement or
analyser
fault is detected.
Therefore if the power to
the analyser
fails or the relay cable is disconnected, then this
is transmitted as a fault, as electrically it has the same effect as if the fault relay is
deenergised, this is often termed "fail safe".
In some applications it is preferred that the fault
relay is energised in an "active"
status, if this is the case then the relay coil set up can be changed as detailed in
5.17.2
Setting up the relay coil
status
To
define the "ac�ve" status of the relay coil, see Sec�on 5.17.1
1.
Select:
Set Up
Relay Set Up
2.
Edit fault relay coil sate that is required for the "ac�ve" status
Figure
52 -
Fault Relay
3.
Edit Maintenance Required relay coil sate that is required for the "ac�ve" status
Содержание SERVOTOUGH Oxy
Страница 2: ...This page inten onally blank...
Страница 4: ...This page inten onally blank...
Страница 20: ...8 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 28: ...16 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 30: ...18 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 41: ...01910001B_8 Servomex Group Limited 2020 29 Figure 12 Fixing dimensions mm...
Страница 82: ...70 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 98: ...86 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 110: ...98 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 112: ...100 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 126: ...114 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 130: ...118 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 132: ...120 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 140: ...128 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 142: ...130 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 144: ...132 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...