95
Fault symptom
Recommended
actions
Analyser
measurements are not as
expected.
Check that the correct display units have been selected,
and that the units conversion factor has been correctly
entered (see Sec�on 5.19).
If you are using pressure compensa�on, check that the
pressure transducer is switched on (see Sec�on 7.2.2).
Ensure there ar
e no leaks in the pipework.
Analyser
measurements are unstable.
Check that the sample gas supply pressure is correct: see
Check that the analyser
is not being subjected to high
levels of vibra�on and ensure there are no leaks.
Check that the sample gas inlet is not blocked, and that
the sample gas supply to the analyser
is not restricted.
Ensure there are no leaks in the pipework.
The
analyser
will not calibrate.
Check that the correct low and high calibra�on gases are
being used: see Sec�on 2.6.
Check that the sample gas inlet is not blocked, and that
the sample gas supply to the analyser
is not restricted.
Check that the sample gas outlet is not blocked, and that
any pipes connected to the outlet are not restricte
d.
The
analyser
will not switch on.
Check that the external supply is switched
on, and that
no fuse or over
-
current device in the external supply has
operated to switch off the supply.
If the external electrical supply is correct, switch off and
isolate the supply and check that the supply is correctly
connected to the analyser
.
If the supply is correctly connected, the opera�ng fuse
may have failed; inspect and replace the fuse if
necessary: refer to Sec�on 8.3.
Содержание SERVOTOUGH Oxy
Страница 2: ...This page inten onally blank...
Страница 4: ...This page inten onally blank...
Страница 20: ...8 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 28: ...16 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 30: ...18 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 41: ...01910001B_8 Servomex Group Limited 2020 29 Figure 12 Fixing dimensions mm...
Страница 82: ...70 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 98: ...86 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 110: ...98 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 112: ...100 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 126: ...114 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 130: ...118 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 132: ...120 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 140: ...128 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 142: ...130 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 144: ...132 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...