![Servomex SERVOTOUGH Oxy Скачать руководство пользователя страница 103](http://html1.mh-extra.com/html/servomex/servotough-oxy/servotough-oxy_operators-manual_1247065103.webp)
01910001B_8
© Servomex Group Limited. 2020
91
Message
Measurement screen icon
Recommended actions
P H/W fault
Contact Servomex or your local
Servomex agent for assistance.
P low fault
Adjust pressure.
Contact
Servomex or your local
Servomex agent for assistance.
Power config fault
Contact Servomex or your local
Servomex
agent for assistance.
Relay not detected
Contact Servomex or your local
Servom
ex agent for assistance.
Relays overridden
Raise
d when the alarm and
status relays have been
overridden (see Sec�on 7.1).
Remote ac�va�o
n
Raised when the Service in
Progress status is remotely
ac�vated via the SIP digital
inputs on the Calibra�on Card.
Rem lo cal denied
Raised when a remote low
calibra�on cannot execute
because the current reading is
outside tolerances (see Sec�on
Rem hi cal denied
Raised when a remot
e high
calibra�on cannot execute
because the current reading is
outside tolerances (see Sec�on
Sta�c RAM fault
Turn the analyser off, and then
turn it on again. If the fault
message is then displayed
again, contact Servomex or your
local Servomex agent for
a
ssistance.
S
w IP not detected
Contact Servomex or your local
Servomex agent for assistance.
Temperature fault
Reduce temperature to within
environmental limits or contact
Servomex or your local
Servomex
agent for assistance.
Содержание SERVOTOUGH Oxy
Страница 2: ...This page inten onally blank...
Страница 4: ...This page inten onally blank...
Страница 20: ...8 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 28: ...16 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 30: ...18 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 41: ...01910001B_8 Servomex Group Limited 2020 29 Figure 12 Fixing dimensions mm...
Страница 82: ...70 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 98: ...86 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 110: ...98 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 112: ...100 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 126: ...114 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 130: ...118 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 132: ...120 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 140: ...128 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 142: ...130 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 144: ...132 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...