![Servomex SERVOTOUGH Oxy Скачать руководство пользователя страница 57](http://html1.mh-extra.com/html/servomex/servotough-oxy/servotough-oxy_operators-manual_1247065057.webp)
01910001B_8
© Servomex Group Limited. 2020
45
5.15.2
Automatic validation/calibration sequence
Before
you carry out an automa�c valida�on/calibra�on, you must
pre-set
the following parameters:
Parameter
Use/range
Low target value
Low measurement target value
for validate/ calibrate.
Low target tolerance
L
ow measurement target tolerance.
High target value
High measurement target value for validate/ calibrate.
High target tolerance
High measurement target tolerance.
Valida�on/Calibra�on type
Low, High, Low & High or High & Low.
Mode
Calibra�on or Valida�on.
Pre-warn
ing
On or Off.
Pre-
warning �me
Only applicable if Pre
-
warning is On. Specifies the length of the pre
-
warning period (0 min 1 sec to 59 min 59 sec)
.
Iner�ng
On or Off.
Iner
t gas
Only applicable if Iner�ng is On. Specifies the type of inert gas (High
or Low).
Iner�ng �me
Only applicable if Iner�ng is ac�ve. Specifies the iner�ng �me (0 min
30 sec to 59 min 59 sec).
Flushing �me
Specifies the flushing �me (0 min 30 sec to 59 min 59 sec).
Timer
On (ac�ve) or Off (inac�ve). If ac�ve the calibra�on or valida�on can
be carried out regularly at a specified frequency.
Start �me/date
Only applicable if the �mer is ac�ve. This specifies the �me/date on
which the first automa�c valida�on/calibra�on will be carried out.
R
epeat �me
Only applicable if the �mer is ac�ve. If set (in the range 1 hour to
999 days), specifies the interval between successive automa�c
calibra�ons/ valida�ons. Select 0 hours and 0 days if repeat is not
required.
shows a typical automa�c calibra�on sequence,
where:
•
Pre-
warning is
on.
•
Automa�c calibra�on has been
ini�ated.
•
Valida�on/Calibra�on type has been set to Low &
High.
•
Mode has been set to
Calibra�on.
•
Iner�ng is set to On
(ac�ve) and Inert gas has been set to Low Calibra�on
Gas.
Содержание SERVOTOUGH Oxy
Страница 2: ...This page inten onally blank...
Страница 4: ...This page inten onally blank...
Страница 20: ...8 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 28: ...16 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 30: ...18 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 41: ...01910001B_8 Servomex Group Limited 2020 29 Figure 12 Fixing dimensions mm...
Страница 82: ...70 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 98: ...86 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 110: ...98 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 112: ...100 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 126: ...114 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 130: ...118 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 132: ...120 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 140: ...128 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 142: ...130 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 144: ...132 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...