OEXP(?)i{,x,j}
The OEXP i command sets or queries the output offsets and expands. The parameter i must
be set when sending this command. The parameter i selects the type of parameters below.
i
Parameter
0
R
1
X
2
Y
3
Rh1
4
Xh1
5
Yh1
6
Rh2
7
Xh2
8
Yh2
9
Noise
The parameter x is used to set the offset and the range is -100.00
≤
x
≤
100.00. The parameter
j is used to set expand and the range is 1
≤
x
≤
256. For example, send command
“
OEXP
2,50.00,2
”
and then the offset of parameter <Y> is 50%; the output expand is 2. Parameter
x and j are required to send.
The OEXP? i command queries the offset and expand of the corresponding parameter of i.
The returned string contains both the offset and expand separated by a comma. For
example, if the OEXP? 2 command returns "50.00,1" then the offset of Y is 50.00% and the
expand of Y is 10.
OAUX (?)i{,x}
The OAUX? Command sets or queries the AUX_DAC output values. The parameter i must
be set when sending this command. The parameter i is used to select a DAC channel and x
is used to set the voltage of DAC. The voltage range is -10.000V≤X≤10.000V. For
example, send command “OAUX 2,5.00” and then the channel voltage of <DAC2> is set to
5.00V.
SPED(?)i
The SPED command sets or queries the executing mode of <Channel Output>. The
parameter i selects <Slow> (i=0) or <Fast> (i=1).
In the <Fast> mode, the output can only choose <R>, <X> and <Y>. If in <Slow> mode
and the current choice is not <R>, <X> or <Y>, then when switching to <Fast> mode, the