<Stop>
stop sweep
<Single>
single sweep
<Loop>
loop sweep
When start from <Single> sweep, SE1022 does an internal amplitude sweep according to the settings of users.
After that, the status jumps back to <Stop> automatically.
When start from <Loop > sweep, SE1022 does constant loop sweep until users switch the status to <Stop>
manually.
3.3 [GAIN/TC]
Press [GAIN/TC] button, then enter the submenu as is shown in Fig.24. There are 5 function settings in the
submenu, including <Sensitivity>, <Reserve>, <Time Constant>, <Filter dB/oct> and <Synchronous>.
Fig.24 [GAIN/TC] Submenu
3.3.1 < Sensitivity>: Setting the Full Scale Sensitivity
Choose an appropriate full-scale sensitivity according to the amplitude of the input signal. Adjust the value through
the soft keys or spin knob.
Table 1. Full Scale Sensitivity
1 nV/fA
200 nV/fA
50 μV/pA
10 mV/nA
2 nV/fA
500 nV/fA
100 μV/pA
20 mV/nA
5 nV/fA
1 μV/pA
200 μV/pA
50 mV/nA
10 nV/fA
2 μV/pA
500 μV/pA
100 mV/nA
20 nV/fA
5 μV/pA
1 mV/nA
200 mV/nA
50 nV/fA
10 μV/pA
2 mV/nA
500 mV/nA
100 nV/fA
20 μV/pA
5 mV/nA
1 V/μA