Coulomb. It's found that from the formula, the contact potential depends on the nature of the metal and the temperature of
the contact surface.
If two dissimilar metals contact, the electromotive force at the contact point will plus a slowly changing millivolt level.
This kind of noise is sensitive to temperature. This noise is typically at very low frequency because the temperature of the
detector and experiment generally changes slowly. This effect is large on the scale of many detector outputs and can be a
problem for low frequency measurements, especially in the mHz range.
Solutions to minimize thermocouple effects:
1) Hold the temperature of the detector or experiment constant.
2) Use a compensation junction.
1.12
Aux In/Out
There are four 16-bit high-precision AUX-ADC input channels in SE1022. The input voltage ranges from - 10 V to + 10
V and the minimum resolution is 0.3 mV. These 4 ADC channels can clamp the input signal and amplify the internal
differential signal. The input impedance is 1 M
Ω
. At the same time, they can pick up signals to measure slow analog
signals or DC signals (the signals come from temperature sensor and pressure sensor). This is useful for proportional
operation and transferring data to computer.
Because of the internal special structure of the 4 AUX-ADC channels, the measurement result is about 2.048V when the
input is floating.
There are also four 16-bit high-precision AUX-ADC output channels in SE1022. The output voltage ranges from - 10 V to
+ 10 V and the minimum resolution is 0.3 mV.
The AUX IN and AUX OUT ports use standard BNC connectors on the rear panel. The setting for AUX ADC/DAC can
be made in the [display] / [AUX OUTPUT] sub-menu.
1.13
Sweep Frequency and Amplitude of Signal Generator
The SE1022 adds some new functions, such as frequency and amplitude scanning. Signal generator can scan the output
amplitude and frequency. Uses can sweep the frequency they interested in and analyze the characteristics of the signal.
This function greatly increased the actual application
The basic way of frequency and amplitude scanning is setting the value of the start and end point, and increasing the value
by step and then the scan is over. See more details in the sections related to frequency and amplitude scanning.
1.14
FFT Spectral Analysis
FFT is Fast Fourier transformation. It is an algorithm that exacts the frequency domain information of the signal.