Tel: 886. 909 602 109 Email: [email protected]
www.salukitec.com
75
Current Impedance 1MOhm Current Gain (when the input signal is small, the gain is 10
6
V/A )
100MOhm Current Gain (when the input signal is small, the gain is 10
8
V/A)
Input Shield
Grounding
Float (the input connector ground and the instrument ground are isolated
by a 10kΩ resistor.)
Ground (t the input connector ground and the instrument ground are
isolated by a 10 Ω resistor.)
Input Coupling
AC coupling
DC coupling
Input Notch Filter
None (Don’t open the notch filter)
Line Notch (Open the 50Hz notch filter to suppress the frequency
interference)
Both Notch Filters (Open the 50Hz and 100Hz notch filters to suppress
the frequency and frequency second harmonic interference).
2x Line Notch (Open the 100Hz notch filters to suppress the frequency
interference).
Note that:
After finishing all the settings, click
to finish the configuration for input signal. All
settings can be changed at the same time and you just need to click this button once.
After changing the settings, you need to click
to finish this configuration. Otherwise the
settings are void.
4. Reference and Frequency Sweep Configuration
The configuration area of reference signal is shown in the red box in Fig.62. The details of configuration are shown in
Table 4.