Press the corresponding soft key and the system will be restarted.
3.10
[AUTO SET]
There are 4 settings in [AUTO SET] submenu, including <Auto Reserve>, <Auto Gain>, <Auto Phase> and < Auto
Scale >, as is shown in Fig.47.
Fig.47 [AUTO SET] Submenu
3.10.1 < Auto Reserve >: Setting the Dynamic Reserve Automatically
Press the < Auto Reserve > button, the system will automatically change the settings of <Reserve> according to the
current signal. The principle is to select the minimum dynamic reserve of the current measurement signal.
3.10.2 < Auto Gain >: Setting the Sensitivity Automatically
Press the < Auto Gain> button, the system will automatically change the setting of <Sensitivity> according to the current
R. The principle is that the system should choose an appropriate <Sensitivity> but no more than the maximum sensitivity.
The < Auto Gain > function in SE1022 needs some response time (generally less than 5s). If the current R fluctuates a
lot, the setting for < Auto Gain > might be fail. In this case, users should switch current menu to the [GAIN/TC]
submenu and set the R manually.
3.10.3 < Auto Phase >: Setting Phase Shift Automatically
Press the < Auto Phase > button and the system will adjust the phase shift of the reference signal to make the phase of
the input signal 0. This < Auto Phase> function needs some response time (generally less than 10s). If the current
θ
fluctuates a lot, the setting for < Auto Phase > might be fail. In this case, users should switch the current menu to
[REF/PHASE] submenu and set
θ
manually.
3.10.4 < Auto Scale >: Setting Scale Automatically