Tel: 886.2.2175 2930 Email: [email protected]
www.salukitec.com
73
c) Number of sweep points: 201
5) Segment sweep setting:
a) Number of open segments: 1
b) Starting frequency: 10MHz
c) Ending frequency: 13.5GHz/26.5GHz/43.5GHz/50GHz/67GHz
d) Number of points: 21
e) Power: -5dBm
f)
Intermediate frequency bandwidth: 1kHz
6) Trigger setting:
a) Trigger source: internal
b) Trigger type: continuous scanning
7) Display format:
Format: logarithmic amplitude
The details of various formats are shown in Table 4.1.
Table 4.1 Default Format Setting of S3602 Series Vector Network Analyzers
Format
Scale
Reference position
Reference value
Logarithmic amplitude
≤10 dB
5
0dB
phase
45
o
5
0
o
Group delay
10ns
5
10fs
Linear amplitude
100mU
5
500mU
Standing wave ratio
1U
5
6U
Real part
2U
5
0U
Imaginary part
2U
5
0U
Polar coordinate
1U
None
1U
Smith chart
1U
None
1U
8) Response setting:
a) Number of channels: 1
b) Intermediate frequency bandwidth: 1kHz
c) Average: closed
d) Average factor: 1
e) Smoothing: closed
f)
Smoothing factor: 2.49% of the value range
g) Electrical delay: 0s
h) Speed factor: 1