Fig. 11.12 Block Diagram I of Connection of Full 2-port Calibration
9) Repeat Step 8 to do the short and Load calibration for Port 1.
10) Repeat Step 8, 9 to do the SOL calibration for Port 2.
11) When single port calibration is done, connect the system as shown in Fig 2. Select [Through] to start through type calibration.
12) Wait for the calibration completed and disconnect the test cable.Now the instrument now is ready to be tested.
Fig. 11.13 Block Diagram II of Connection of Full 2-port Calibration
Marker Operations
The measurement will need to use Marker operations.
When press [Marker] key, a soft menu will pop out. Then select [Marker Search], the marker search dialog pops out as shown in
figure below. In this dialog, user can select a marker, activate the search function (Max, Min etc.) in a configurable frequency span
(Full Span /User Span)