Tel: 886.2.2175 2930 Email: [email protected]
www.salukitec.com
138
5.2.6.3.
Interpolation ON/OFF
Function description:
Select this menu to enable the interpolation function.
Parameter description:
OFF [OFF|ON].
5.2.6.4.
Port Extension
Function description:
Select this menu to move the measurement reference plane in an electrical manner. Thus, other kinds of calibration can be avoided.
Add a cable after calibration. The port extension function is used to “tell" the length of the cable added to the specific port of the analyzer.
In case of failure of direct calibration, the delay (phase shift) caused by the clamp can be compensated by port extension. For the method
of port extension setting, refer to the section of port extension in 5.5 Improvement of Phase Measurement Accuracy.
Parameter description:
None
5.2.6.5.
Fixture>>
This menu is used to set the fixture function. Click it to enter the menu of lower level. The specific menu includes:
Table 5.18 Fixture
Menu
Fixture ON/OFF
Port Matching...
2-port de-embedding
Port impedance conversion...
Four-port fixture de-embedding/embedding
1) Fixture ON/OFF
Function description:
Select this menu to enable the fixture simulator function. The functions of the fixture simulator are effective only in this mode.
Parameter description:
OFF [OFF|ON].
2) Port Matching...
Function description:
Select this menu to enable the Port Matching... function. For setting of Port Matching..., refer to “7.9.3 Port Matching...” in “7.9 Fixture
Compensation Calibration”.