5) Click [Analysis] > [trace statistics] > [trace statistics ON/OFF].
Fig. 11.7 Block Diagram of System Trace Noise Test
11.2.10.4.
Measurement of S12 amplitude trace noise
1) Click [Measurement] > [S12].
2) Record the deviation displayed on the screen after sweep of the vector network analyzer.
11.2.10.5.
Measurement of S12 phase trace noise
1) Click [Response] > [Format] > [Phase].
2) Record the deviation displayed on the screen.
11.2.10.6.
Measurement of S21 phase trace noise
1) Click [Measurement] > [S21].
2) Record the deviation displayed on the screen after sweep of the vector network analyzer.
11.2.10.7.
Measurement of S21 amplitude trace noise
1) Click [Response] > [Format] > [Logarithmic amplitude].
2) Record the deviation displayed on the screen.
3) Change the frequency value and repeat Step (4) to (13). Measure the trace noise value of other frequency points.
11.2.11.
Dynamic Range Test of System
11.2.11.1.
Items of Dynamic Range Test of System
Test the indicators of the dynamic range of the system after response isolation calibration with S3602 series vector network
analyzers.
11.2.11.2.
Instrument and Accessories for Dynamic Range Test of System
1)
3.5mm coaxial calibration kit
SAV31121 (SAV3602A/B) or SAV31123 (AV3602C/D)
2)
Soft test cable
Agilent 85131F (AV3602A/B) or FE0BN0BM025.0 (AV3602C/D)
11.2.11.3.
Steps of Dynamic Range Test of System
Test of dynamic range of Port 1:
1) As shown in Fig. 11.8, connect the system and preheat is for 60min at least.
2) Click [Reset] and set the S3602 series vector network analyzer into the initial state. Click [Power] > [+20] and press the
ENTER key. Click [Channel] > [Hardware configuration] >[Path configuration]. Select the high power mode for the low-band
segment of Source 1 and 3.