Tel: 886.2.2175 2930 Email: [email protected]
www.salukitec.com
260
Fig. 9.5 LAN Interface of Rear Panel of Instrument
9.2.1.2.
Interface Configuration
Ensure that the physical connection is smooth for remote control of the vector network analyzer by LAN. As the DHCP, domain name
access and WAN are not supported, the network programming of the vector network analyzer is relatively simple. In the menu shown
in Fig. 9.6, set the IP address", Subnet mask" and Default gateway" to be within the subnet with the master controller.
Ensure normal physical connection of the vector network analyzer through the 10Base-TLAN or 100Base- TLAN cable.
The vector network analyzer only supports the establishment of one LAN control system and static IP address setting, and does not
support DHCP or access to the host through the DNS and domain name server. Therefore, it is not required to modify the subnet mask.
The subnet mask of the instrument should be set as the fixed value 255.255.255.0.
9.2.2.
GPIB
9.2.2.1.
Establish Connection
Connect the S3602 series vector network analyzer to the external controller (computer) through the GPIB cable, as shown in Fig. 9.6.
Fig. 9.6 Connection Diagram of GPIB Interface
Fig. 9.7 GPIB Interface of Rear Panel of Instrument