Fig. 7.15 [Open-circuit device] Dialog Box
1) Characteristic zone
a) [Standard calibration kit ID] box
Display or edit the serial number of the standard calibration kit.
b) [Name] box
Display or edit the name of the standard calibration kit.
c) [Description] box
Display or edit the characteristic description of the standard calibration kit.
2) Frequency range
a) [Minimum frequency] box
Define the minimum working frequency of the open-circuit device in calibration.
b) [Maximum frequency] box
Define the maximum working frequency of the open-circuit device in calibration.
3) Connector zone
Define the connector polarity of the open-circuit device: male, female or no gender.
4) Open-circuit characteristic zone
Define the fringe capacitance of the open-circuit device with [C0], [C1], [C2] or [C3].
5) Delay characteristic zone
a) [Delay] box
Define the one-way transmission time from the calibration plane to standard calibration kit.
b) [Loss] box
Define the energy loss caused by the skin effect of the standard calibration kit, in ohms/S, corresponding to the frequency in 1GHz.
Respectively measure the loss and delay time of the standard calibration kit at 1GHz and calculate the loss by the following formula: