e) Measurement application:
Measurement of all S-parameters.
Improvement of the measurement accuracy by 12 error corrections.
7.3.
Calibration Guide
Select the calibration type and perform calibration according to the calibration guide of the analyzer. The through type response and
isolation calibration are taken as examples below to demonstrate the calibration process (this kind of calibration is effective only in
transmission measurement).
Enable the calibration guide for calibration.
1) Menu path: [Calibration] > [Calibration…]. The [Calibration guide] dialog box will appear.
Fig. 7.1 Enabling of Calibration
2) Click [Unguided calibration] and [Next].
Fig. 7.2 Start of Calibration
3) Click [Through type response and isolation calibration] and [Next].
Fig. 7.3 Selection of Type of Mechanical Calibration
4) Click [Select calibration kit], and the [Select calibration kit] dialog box will appear. Select the calibration kit and click [OK] to open
the dialog box. Thus the calibration kit is changed.