22
3) Electrostatic protection
Static electricity is destructive to electronic components and devices. Generally, two anti-static measures can be taken: combination of
conductive table mat and wrist strap; and combination of conductive floor mat and ankle strap. Anti-static effect will be better when two
measures are used at the same time. If used independently, only the former can make sure the safety. In order to ensure the user safety,
the ground isolation resistance of anti-static components must be 1MΩ at least.
The follow anti-static measures should be taken to reduce static damage:
All instruments must be properly grounded to prevent static electricity.
The internal and external conductor of the cable should respectively contact with the ground temporarily before the coaxial cable
and instrument are connected.
Workers must wear the anti-static wrist strap or take other anti-static measures before contact with the connector or wire or any
assembly.
Voltage range
The above anti-static measures are not suitable for the places with the voltage higher than 500V.
3.1.1.3.
On/Off
1) Precautions before powering-on
Check the following items before powering on the instrument.
a. Confirm the power supply parameters.
S3602 series vector network analyzers are equipped with a three-core power line interface and conform to the international safety
standards. Confirm that the protective ground wire of the power supply socket is grounded reliably before inserting the power line into the
standard three-core socket to power on the instrument. Floating or improper grounding may result in damage to the instrument and even
personal injuries. Do not use the power line with no protective measure. Table 3.4 lists the requirements of the external power supply for
normal operation of the network analyzer.
Table 3.4 Parameters of Working Power Supply of S3602
Power supply parameter
Applicable range
Voltage,frequency
220V±10% and 50-60Hz
Rated output current
>3A
Power consumption
>600W
Prevent mutual interference of various power supplies.
In order to prevent mutual interference of power supplies of various devices, especially the instrument hardware damage caused by peak
pulse interference of high-power equipment, it is recommended to use the 220V AC voltage-regulated power supply to power on the
network analyzer.