Chapter 6: Arrhythmia Monitoring
User Manual
6-7
●
RATE
With count, you can determine the point at which an event call is triggered.
You canno
t modify the rate for “
ASYSTOLE
”, ”
VFIB
”, ”
COUP
LET”, ”
BIGEM
INY”,
”TRIGEMINY”, “PAUS”, “ AFIB” and “FREQUENT PVCs” .
“RUN” and “AIVR” derive their
rate
settings from “VTAC” and cannot be
modified.
Rate setting
Arrhythmia event
100-200 step by 10
VTAC
Same as VTAC rate
RUN
<VTAC rate-1
AIVR
100-200 step by 10
TACHY
30-105 step by 5
BRADY
●
COUNT
With rate, you can determine the point at which an event call is triggered.
You can’t modify the count for “
ASYSTOLE
”, ”
VFIB
”, ”
COUP
LET”, ”
BIGEM
INY”,
”TRIGEMINY”,”TACHY”,”BRADY”, “AFIB” and “PAUS”.
Count of “AIVR” is ≥3
and cannot be modified.
Count setting
Arrhythmia event
5-12
step by 1
VTAC
(VTACcount -1) ~3 step by
1
RUN
1-15 step by 1
FREQUENT PVCs
●
ARCHIVE
You can determine whether the selected event is stored, recorded automatically or both.
You can view stored events on ARR EVENT RECALL Window.
STR: Stores selected arrhythmia event.
REC: Automatically generates a recording of selected event.
STR/REC: Event is stored and recorded simultaneously.
OFF: No action if arrhythmia event activates.
●
ALL ALM LEVEL
Press to set the level of all arrhythmia alarms to the same value or to disable all of them.
●
ALL ARCHIVE
Press to set all arrhythmia ARCHIVE condition to the same state.
Содержание ALBORZ B9