Chapter 2: Configuration
User Manual
2-4
When Display Format is set to7 Traces, seven traces of ECG signal are displayed).
P3
:
You can monitor ECG (2 traces), CO2 (RESP) and SpO2 signals as well as all numeric
parameters (special page for Rainbow parameters) (MAIN DISPLAY= PAGE 3 mode)
P4:
You can monitor ECG signal (12 traces) and numeric parameters of P1.
(Only with settings of CABLE TYPE: 10WIRES and DISPLAY FORMAT: 12 TRACES,
you can access this page)
If IBP module is ON:
NOTE:
In all pages except page 6, you can monitor BFA signal and numeric parameters only if
BFA module is enabled.
P1:
The following signals will be displayed in this page for different Display Formats:
a) CASCADE/2 TRACES: ECG signal (two traces), SpO2, IBP1, IBP2 and RESP/CO2
b) 4 TRACES: ECG (four traces), IBP1 and IBP2 signals
c) 7 TRACES: Seven traces of ECG signal
Numeric parameters HR, PVCs, ST, SpO2%, PR, IBP1, IBP2, RESP/CO2, NIBP and TEMP
are also displayed in this page. (MAIN DISPLAY= PAGE 1 mode)
P2
:
You can monitor same parameters and signals as P1, but arrangement of parameters
windows differ slightly from P1. (MAIN DISPLAY= PAGE 2 mode)
(When Display Format is set to 4 Traces, four traces of ECG signal as well as SpO2 and
RESP/GAS signals are displayed in this page.
When Display Format is set to7 Traces, seven traces of ECG signal are displayed).
P3
*
:
You can monitor ECG (2 traces), CO2 (RESP), IBP1/IBP2 and SpO2 signals as well as
all numeric parameters (special page for Rainbow parameters) (MAIN DISPLAY= PAGE 3
mode).
P4
: You can monitor ECG, IBP1 and IBP2 signals (in a larger scale) as well as numeric
parameters HR, PVCs, ST, SpO2%, PR, IBP1,IBP2, NIBP and TEMP (special page for large
scale IBP) (MAIN DISPLAY= PAGE 4 mode)
P5:
You can monitor ECG (two traces), IBP1, SpO2 and RESP signals as well as all numeric
parameters of P1 except IBP2 (MAIN DISPLAY= PAGE 5 mode).
P6
*
:
You can monitor ECG, SpO2, RESP/CO2 and two IBP (IBP1,2 or IBP 3,4) signals as
well as all numeric parameters of P1 and IBP3,4 parameters. (special page for 4 channels
Содержание ALBORZ B9
Страница 1: ...POOYANDEGAN RAH SAADAT OPERATOR S MANUAL ALBORZ B9 Patient Monitor D00132 15...
Страница 153: ...Chapter 12 IBP Monitoring User Manual 12 14 b IBP WINDOW...
Страница 240: ...ALARM RECALL User Manual 17 11 Recording alarm list Click on RECORD in HOME ALARM RECALL WINDOW to record the alarm list...