Chapter 5: ECG Monitoring
User Manual
5-25
ECG
CHECK LA
OR ALL
LA or other leads are
not properly
connected when ECG
lead is II for 3wire
lead set
Make sure that all electrodes and
patient cable are properly
connected
Alarm level 2- the message is
displayed in yellow background. By
pressing ALARM SILENCE,
background becomes gray and the
alarm is disabled and ignores this
fault.
ECG
CHECK RA
OR ALL
RA or other leads are
not properly
connected when ECG
lead is III for 3wire
lead set
Make sure that all electrodes and
patient cable are properly
connected
Alarm level 2- the message is
displayed in yellow background. By
pressing ALARM SILENCE,
background becomes gray and the
alarm is disabled and ignores this
fault.
ECG
CHECK C
(C2, C3, C4,
C5, C6 )
C lead is not properly
connected to the
patient.
Make sure that all electrodes esp.
C and ECG cable are properly
connected.
Alarm level 2-the message is
displayed in yellow background. By
pressing ALARM SILENCE,
background becomes gray and the
alarm is disabled and ignores this
fault.
After checking the mentioned solution, if the alarm persists, the ECG cable may be damaged and
you should contact with local After Sales Services.