Chapter 13: GAS Monitoring (Mainstream)
User Manual
13-12
The Capnography window for Mainstream sensor in different modes is as follows:
a) CO2 WINDOW in CO2 (ONLY) mode
b) GAS WINDOW in AX+ mode
Figure 10-5 Capnography window of Mainstream sensor in different modes
■
Co2 UNIT
Pick this item to adjust measurement unit. (Options: mmHg, KPa, %V)
EtCo
2
in %V is the Co
2
value (in mmHg) divided by ambient barometric pressure (in
mmHg) which is a percentage of the barometric pressure.
)
(
)
(
2
mmHg
Brometric
mmHg
EtCo
p
P
EtCo2(%V)
=
Содержание ALBORZ B9