Chapter 16: Cardiac Output Monitoring
User Manual
16-9
Measurement Mode
To start measurement in MANUAL mode, press START key in CARDIAC OUTPUT MENU
after you prepared injectate solution.
Catheter Type
To select catheter type. Available options are
“
131HF7
”
and
“
139HF75P
”
Edwards catheters
and “
Simulator
”
. The last option is intended to use simulator with Comp.Con. of 0.542.
Comp. Con.
The computation constant changes based on catheter type. Make sure that appropriate catheter
is selected, because computation constant directly influences C.O. measurement.
C.O. settings in SETUP menu can be monitored in CARDIAC OUTPUT MENU to check and
modify any possible error occurred in data input.