APPENDIX II
User Manual
Appendix II-1
APPENDIX II
EMC
Warning
Use only the recommended manufacturer accessory .Using the accessory other than in
relevant chapter may cause to increase the EMISSION or decrease the IMMUNITY of
system.
Warning
Measurements can be affected by mobile and RF communications equipment. It should be
assured that the bedside monitor is used in the electromagnetic environment specified.
Warning
To prevent EMC effect on the monitor, the system should not be used adjacent to or
stacked with other equipment and that if adjacent or stacked use is necessary, the
equipment should be observed to verify normal operation in the configuration in which it
will be used.
Warning
Do not use cellular phone in the vicinity of this equipment. High level of electromagnetic
radiation emitted from such devices may result in strong interference with the monitor
performance.
Guidance and manufacturer's declaration
–
electromagnetic emissions
The patient care monitors are intended for use in the electromagnetic environment specified below.
The customer or the user of the monitor should assure that they are used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The use RF energy only for their internal function.
Therefore, their RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
The monitors are suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Complies
Voltage fluctuations/ flicker
emissions
IEC 61000-3-3
Complies