Chapter 12: IBP Monitoring
User Manual
12-17
c) Messages
IBP MESSAGES
Message
Cause
Solution
Explanation
IBP1/IBP2
ADJUST
SCALE
IBP1 or IBP2 signal is out of display
range for about 5 seconds.
Press <AUTO SCALE> in
IBP WINDOW.
IBP1/IBP2
SEARCH
IBP signal can’t be processed by the
software because the signal is weak or
less pulsatile.
●
Check all IBP measurement
setup is suitable or not.
●
Check patient status and
treat him, if necessary
Содержание ALBORZ B9